8 Specifications

30

Expected Sensor Life
Considers the full range of the sensor’s signal, example 7-13 mV. Most medical
oxygen sensors are configured to meet the published, see opposite page, speci-
fication which distributes the overall sensor life as follows:

- 60 months Expected Service Life (915,420 oxygen % hours)
- 6 months Recommended Storage Life period (91,542 % oxygen hours)
- 2 months margin of error


Therefore, the Recommended Storage life period should not be considered a
perishable shelf life. Operating at the specified parameters of oxygen concen-
tration (air 20.9%), temperature (25⁰C/77⁰F) and pressure (1 atm/bar), the
sensor will operate for approximately 68 months whether in storage or in use.

The purpose of the Storage Life period is to ensure the user derives the Ex-
pected Life of 60 months (915,420 % oxygen hours) and does not lose the
benefit of the warranty.

Warranty
The 18 month (274,626 % oxygen hours) warranty period (begins with ship-
ment from the factory and is limited to the first claim submitted) is based on:

- 60 months Expected Service Life (915,420 % oxygen hours)
- Estimated exposure (24/7) to 60-70% oxygen concentration
- Marginal of error of 2 months

8.1 Spare Parts & Accessories

31

Accuracy: +2% of FS range under constant conditions

Analysis: 0-100%

oxygen

Area Classification: General purpose

Alarms:

Analyzer – none; Monitor - User adjustable HI 16-100%,
LO 15-99%, 120 second alarm silence, HI alarm defeat
for flushing patients with 100% O

2

Calibration:

100% oxygen before using each day or after 8 hours of
continuous use.

Compensation: Temperature

compensated

Connections:

(M, A) 1x16mm thread or o-ring diverter; (HC) tubing

Controls:

Soft touch keypad for ON/OFF and menu function

Dimensions:

3.6 x 5.9 x 1.6”; weight 10 oz. (280 grams)

Display:

3-1/2 digit backlit LCD 2.5” x 1.5”; resolution 0.1% O

2

Flow Sensitivity:

None between 0.2 to 10 liters per minute

Humidity: Non-condensing

0-95%

RH

LED Indicators:

Analyzer - none; Monitor - upon activation of alarms

Linearity: + 1% under constant conditions

Pressure:

Inlet – (M, A) ambient, (HC) regulate; vent - atmospheric

Power:

2 AA Alkaline batteries; 1,200 hours continuous use

Response Time:

90% of final FS reading in 9 seconds

Sensitivity:

< 0.5% of FS range

Sensor:

(M, A) AII-11-60 or (HC) AII-11-60-HC

Expected Life:

60 months in air at 25ºC and 1 atmosphere

Storage Temp.:

-20º to 60ºC (-4ºF to 140ºF) on intermittent basis

Temp. Range:

5º to 45ºC (41ºF to 113ºF)

Warm-up Time:

None

Warranty:

24 months analyzer; 18 months sensor (any application)

AII-2000 A, AII-2000 M AII-2000

HC

Spare Parts:

Spare Parts:

AII-11-60 Oxygen Sensor

AII-11-60-HC Oxygen Sensor

BATT-1008 Battery (2x) 1.5V AA

BATT-1008 Battery (2x) 1.5V AA

P-1087 Instructions for Use

P-1087 Instructions for Use

A--1162 PCB Assy Main

A-1162 PCB Assy Main

CABL-1006 Coil Cable

TUBE-1007 1/4” Tubing 7’

FITN-1112-1 Flow Diverter

Optional Accessories:

Optional Accessories:

FITN-1009 Tee Adapter

FITN-1066 Nipple Universal

HRWR-1075 Dovetail Clamp

HRWR-1074 Dovetail Clamp

CC-1072 Carrying Case

CC-1072 Carrying Case