4.5 Start-Up Test
Press the ON/OFF key on the front panel to apply power to the device and
initiate a complete diagnostic test of all system functions: the electronics,
feeds voltage and tests the alarm circuit (monitor only below right) internally,
confirms the battery voltage is adequate to power the circuit, and, the sen-
sor’s signal output is within specifications.









Following successful Start-Up Test the devices default to the SAMPLING mode.








With the exception of the ALARMS for the AII-2000 M (above left) the tests
and resulting displays are the same.


Note: Any START-UP TEST failure requires the user to take corrective
action before continuing or attempting to use any device.


4.5.1 Electronics, Alarms (AII-2000 M Monitor) or Battery Failure
If any of these START-UP TESTs are unsuccessful, the following display in-
structs the user to SERVICE DEVICE. The following display is the same for all
models.

12

4.5.2 Sensor Failure
Sensor failure can result from multiple causes; the user’s failure to connect a
sensor or sensor cable, a defective sensor cable or a sensor with an output
outside specification.

SENSOR - FAILED LOW is one of the possible unsuccessful START-UP TESTs as
illustrated previously and displays additional warnings as follows.

4.5.2.1 AII 2000 A and AII 2000 HC Oxygen Analyzers
The LCD alternately displays the following until the problem is corrected.








Corrective action:
1. Expose the sensor to air or a gas containing approximately 20.9% oxygen
2. Connect or replace the cable connecting the sensor to the analyzer
3. Connect or replace the oxygen sensor

4.5.2.2 AII 2000 M Oxygen Monitor
Performs the same routine and requires the same corrective action as the ana-
lyzers above with additional indicators related to the monitor’s alarm feature.








In addition to the alternating LCD display, the LO ALARM becomes active and:

LO ALARM value and background alternately reverse colors on the LCD

RED LED below the LO ALARM value lights up and begins flashing

Audible alarm begins beeping


The audible alarm can be disabled for two (2) minutes (unlimited times) by:
1. Press the MENU key on the front panel
2. Press the UP/DOWN arrow to select ALARMS AUDIBLE
3. Press the ENTER key to toggle to ALARMS SILENT mode

13

START-UP TEST

ELECTRONICS - PASS

ALARMS - N/A

BATTERY - PASS

SENSOR - PASS

START-UP TEST

ELECTRONICS - PASS

ALARMS - PASS

BATTERY - PASS

SENSOR - PASS

20.9 %

SAMPLING

LO 15% HI 50%

20.9 %

SAMPLING

START-UP TEST

ELECTRONICS - FAILED

ALARMS - FAILED

BATTERY - FAILED

SENSOR - FAILED LOW

SERVICE DEVICE

0.0 %

ALARM

0.0 %

ALARM

LO 15% HI 50%

CHECK SAMPLE GAS

CHECK CABLE

CHECK SENSOR

ALARM

CHECK SAMPLE GAS

CHECK CABLE

CHECK SENSOR

ALARM

LO 15% HI 50%