Atec Agilent-N8300A User Manual
Page 12
12
Amplitude specifications for the RF Output port
Output amplitude range
1
2.402 to 2.484 GHz: +20 to -100 dBm (CW);
4.8 to 5.875 GHz: +13 to -100 dBm (CW);
See EVM specification for information on modulated
signals
Absolute amplitude
accuracy
2.402 to 2.484 GHz:
±0.9 dB (CW) (+7 to -90 dBm)
±0.5 dB (CW) typical (+7 to -90 dBm)
4.8 to 5.875 GHz:
±1.0 dB (CW) (+7 to -90 dBm)
±0.5 dB (CW) typical (+7 to -90 dBm)
Error vector magnitude
(EVM)
802.11a,g,n: 2.4 GHz band; 54 Mbps 64QAM OFDM:
-44 dB (<0.6%), up to +6 dBm
802.11a,g,n: 5 GHz band; 54 Mbps 64QAM OFDM:
-44 dB (<0.6%), up to +2 dBm
802.11b: DSSS: <-36 dB (< 1.5%), up to +6 dBm
Spurious (non-harmonics)
2.402 to 2.484 GHz: < -62 dBc typical (in-band)
4.8 to 5.875 GHz: < -56 dBc typical (in-band)
1
This specification applies to instruments serial prefix MY4832 or greater, otherwise the specification is 2.402 to
2.484 GHz:+10 to -100 dBm (CW); 4.8 to 5.875 GHz: +6 to -100 dBm (CW)