Required tools – Atec Agilent-85133F User Manual

Page 120

Advertising
background image

124

Installation and User’s Guide

C

RF Subsystem Functional Verification Test

Understanding the RF Subsystem Functional Verification Test

Use the procedure provided in

“Performing the RF Subsystem Functional

Verification Test

to manually confirm the functionality of the

Agilent E8364B PNA Series vector network analyzer system. This
procedure does not verify performance to specification. This procedure
confirms that the network analyzer is ready for performance verification
and/or operation by confirming the following hardware functionality:

the repeatability of the RF switch in the test set

the attenuation range of the test port attenuators

Required Tools

Agilent E8364B PNA Series vector network analyzer

Test port cable

a standard short from the network analyzer calibration kit