S-parameter measurement, Corrected system performance, Description specification spd – Atec Agilent-E5061B User Manual

Page 3

Advertising
background image

3

S-Parameter Measurement

Corrected system performance

The specifications in this section apply for measurements made with the Agilent
E5061B network analyzer with the following conditions:
• No averaging applied to data

• Environmental temperature of 23 °C ±5 °C, with less than 1 °C deviation from

the calibration temperature

• Response and isolation calibration not omitted

Table 1. System dynamic range

1,2

Description

Specification SPD

System dynamic range

(Option 3L5)

100 kHz to 1 MHz, 3 kHz IF bandwidth

90 dB

1 MHz to 3 GHz, 3 kHz IF bandwidth

95 dB

5 to 100 Hz, 2Hz IF bandwidth

90 dB

100 Hz to 9 kHz,10 Hz IF bandwidth

100 dB

9 to 100 kHz, 10 Hz IF bandwidth

110 dB

100 kHz to 1 MHz, 10 Hz IF bandwidth

115 dB

1 MHz to 3 GHz, 10 Hz IF bandwidth

120 dB

130 dB

(Option 115, 135, 215, 235, 117, 137, 217, 237)

100 to 300 kHz, 3 kHz IF bandwidth

75 dB

300 kHz to 1 MHz, 3 kHz IF bandwidth

90 dB

1 MHz to 3 GHz, 3 kHz IF bandwidth

95 dB

100 to 300 kHz, 10 Hz IF bandwidth

100 dB

300 kHz to 1 MHz, 10 Hz IF bandwidth

115 dB

1 MHz to 3 GHz, 10 Hz IF bandwidth

120 dB

130 dB

1. The test port dynamic range is calculated as the difference between the test port rms noise fl oor

and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.

2. The specifi cation might not be met at the frequencies 1.4 MHz, 4.0 MHz, 4.333 MHz, 6.333

MHz, 25MHz and 90 MHz.