Table 6. Corrected system performance with Type-N 50 Ω connectors,
85032F calibration kit, enhanced response calibration

Network analyzer: E5061B-115/135/215/235, calibration kit: 85032F (Type-N, 50 Ω),
calibration: enhanced response calibration

IF bandwidth = 10 Hz, No averaging applied to data, environmental temperature
= 23 °C ±5 °C with < 1 °C deviation from calibration temperature, isolation calibration
not omitted

Description

Specification (dB)

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

Directivity

49

49

46

Source match

41

41

40

Load match

48

49

46

Refl ection tracking

0.011

0.011

0.021

Transmission tracking

0.035

0.028

0.034

0.01

0.1

1

10

-90

-80

-70

-60

-50

-40

-30

-20

-10

10

0

Uncertainty (dB)

Transmission Coefficient (dB)

Magnitude

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

0

0.01

0.02

0.03

0.04

0.05

0

0.2

0.4

0.6

0.8

1

U

n

cer

tai

n

ty (

Li

n

ear

)

Reflection Coefficient (Linear)

Magnitude

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

0

2

4

6

8

10

0

0.2

0.4

0.6

0.8

1

Uncertainty (Degrees)

Reflection Coefficient (Linear)

Phase

100 kHz to 300 kHz

300 kHz to 1 MHz

1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

0.1

1

10

100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Uncertainty (Degrees)

Transmission Coefficient (dB)

Phase

100 kHz to 300 kHz

300 kHz to 1 MHz
1 MHz to 3 GHz

S11 = S22 = 0; Source Power = -10 dBm

Figure 9. Transmission uncertainty (specifi cation)

Figure 10. Refl ection uncertainty (specifi cation)