Atec Agilent-E5100A User Manual

Agilent e5100a/b network analyzer, Data sheet

Advertising
background image

These specifications are the performance standards or limits against which the instrument is tested. When shipped
from the factory, the E5100A/B meets the specifications listed in this section.

Values followed by (SPC) are supplemental performance characteristics.

Source

Frequency characteristics

Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 kHz to 300 MHz
Accuracy (at 23 ±5°C) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±20 ppm

with Option 1D5 (at 0 to 55°C, 20 minutes after power on) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 ppm

Stability (at 23 ±5°C) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±5 x 10

–6

/day (SPC)

with Option E5100x-1D5 (48 hours after power on) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±2.5 x 10

–9

/8 hours (SPC)

Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 mHz

Output power characteristics
(measured at RF OUT 1, RF OUT 2 is terminated with 50

termination)

Range (nominal)

with Option E5100x-001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –9 dBm to +11 dBm
with Option E5100x-002 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –15 dBm to +5 dBm
with Option E5100x-003 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –12 dBm to +8 dBm
with Option E5100x-801 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –48 dBm to +22 dBm
with Option E5100x-802 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –54 dBm to +16 dBm
with Option E5100x-803 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –51 dBm to +19 dBm
with Option E5100x-600 (at RF OUT 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –52 dBm to +18 dBm
with Option E5100x-600 (at RF OUT 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –65 dBm to +5 dBm

Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1 dB
Level accuracy (at 23 ±5°C, 0 dBm output level, 50 MHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
Flatness (at 23 ±5°C, relative to 0 dBm output level at 50 MHz) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +2 dB, –4 dB

with Option E5100x-803 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +2.5 dB, –4.5 dB

Agilent E5100A/B

Network Analyzer

Data Sheet

Option ordering information: To add options to a product, use the following ordering scheme:

Model: E5100x, where x = A or B
Example model option: E5100x-Option 001

Note: Options E5100x-600/704/706/708 are for E5100A only.

Advertising