Table 1-14

Test port input (group delay

1

)

Description Specification

Supplemental

information

Aperture (selectable)

(frequency span)/(number of points -1)

Maximum aperture

25% of frequency span

Maximum delay

Limited to measuring no more than

180° of phase change within the minimum aperture.

Accuracy

See graph below, typical

The following graph shows group delay accuracy with type-N full 2-port calibration and a 10 Hz IF bandwidth. Insertion loss is assumed to be < 2 dB.

Group delay (typical)

In general, the following formula can be used to determine the accuracy, in seconds, of specific group delay measurement:
± phase accuracy (deg)/[360 x aperture (Hz)]

1

Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep).