Corrected system performance – Atec Agilent-E5071B User Manual

Page 3

Advertising
background image

3

Corrected system performance

The specifications in this section apply for measurements
made with the Agilent E5070B/E5071B network analyzer
with the following conditions:

• No averaging applied to data
• Environmental temperature of 23°C ± 5°C, with less
than 1°C deviation from the calibration temperature
• Response and isolation calibration not omitted

Table 1-1

System dynamic range

1,2

Description Specification

Supplemental

information

System dynamic range

300 kHz to 3 MHz, IF bandwidth = 3 kHz

85 dB

3 MHz to 1.5 GHz, IF bandwidth = 3 kHz

95 dB

98 dB

1.5 GHz to 3 GHz, IF bandwidth = 3 kHz

97 dB

100 dB

3 GHz to 4 GHz, IF bandwidth = 3 kHz

96 dB

99 dB

4 GHz to 6 GHz, IF bandwidth = 3 kHz

92 dB

94 dB

6 GHz to 7.5 GHz, IF bandwidth = 3 kHz

87 dB

90 dB

7.5 GHz to 8.5 GHz, IF bandwidth = 3 kHz

80 dB

83 dB

300 kHz to 3 MHz, IF bandwidth = 10 Hz

110 dB

3 MHz to 1.5 GHz, IF bandwidth = 10 Hz

120 dB

123 dB

1.5 GHz to 3 GHz, IF bandwidth = 10 Hz

122 dB

125 dB

3 GHz to 4 GHz, IF bandwidth = 10 kHz

121 dB

124 dB

4 GHz to 6 GHz, IF bandwidth = 10 Hz

117 dB

119 dB

6 GHz to 7.5 GHz, IF bandwidth = 10 Hz

112 dB

115 dB

7.5 GHz to 8.5 GHz, IF bandwidth = 10 Hz

105 dB

108 dB

1

The test port dynamic range is calculated as the difference between the test port rms noise floor

and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.

2

May be limited to 90 dB at particular frequencies below 350 MHz or above 4.25 GHz due to

spurious receiver residuals.

Advertising