General specifications

Power supply

Power adapter

Line voltage range: 50/60 Hz, 100 to 240 VAC, 1.6 A

Output voltage: 15 VDC, 4 A

Installation Category II

Battery

Li-Ion rechargeable battery pack, 10.8 V

Operating time: Up to 3 hours

Operating environment

Temperature

0 to 50 °C (with battery only)

0 to 40 °C (with power adapter)

Humidity

0 to 80% RH (0 to 35 °C)

0 to 50% RH (35 to 40/50 °C)

Altitude up to 2000 m

Pollution degree 2

Storage compliance

Temperature

-20 to 70 °C

Humidity

0 to 80% RH

Altitude up to 15000 m

Shock

Tested to IEC 60068-2-27

Vibration

Tested to IEC 60068-2-6, IEC 60068-2-64

Safety compliance

IEC 61010-1:2001/EN 61010-1:2001

Canada: CAN/CSA-C22.2 No. 61010-1-04

USA: ANSI/UL 61010-1:2004

EMC compliance

IEC 61326-1:2005/EN 61326-1:2006

Australia/New Zealand: AS/NZS CISPR 11:2004

Canada: ICES/NMB-001:ISSUE 4, June 2006

IP rating

IP 41 ingress protection according to IEC 60529

Dimensions (W × H × D)

183 x 270 x 65 mm

Weight

< 2.5 kg