N1923A/N1924A Wideband Power Sensor Specifications

Sensor
model

Frequency
range

Dynamic range

Rise/fall time

Damage level

Connector
type

N1923A 50 MHz to

18 GHz

–35 dBm to +20 dBm

≤ 3 ns (applicable for
frequencies of ≥ 500
MHz)

+23 dBm (average power);
+30 dBm (< 1 μs duration,
peak power)

Type N (m)

N1924A 50 MHz to

40 GHz

–35 dBm to +20 dBm

≤ 3 ns (applicable for
frequencies of ≥ 500
MHz)

+23 dBm (average power);
+30 dBm (< 1 μs duration,
peak power)

2.4 mm (m)

The N1921A/N1922A P-series wideband power sensors are compatible for use with the 8990B peak power analyzer.

Maximum SWR

Frequency band

N1923A

N1924A

50 MHz to 10 GHz

1.2

1.2

10 GHz to 18 GHz

1.26

1.26

18 GHz to 26.5 GHz

1.3

26.5 GHz to 40 GHz

1.5

Sensor Calibration Uncertainty

1

Frequency band

N1923A

N1924A

50 MHz to 500 MHz

4.5%

4.3%

500 MHz to 1 GHz

4.0%

4.2%

1GHz to 10 GHz

4.0%

4.4%

10 GHz to 18 GHz

5.0%

4.7%

18 GHz to 26.5 GHz

5.9%

26.5 GHz to 40 GHz

6.0%

Physical characteristics

Dimensions

N1923A
N1924A

135 mm x 40 mm x 27 mm (5.3 in x 1.6 in x 1.1 in)
127 mm x 40 mm x 27 mm (5.0 in x 1.6 in x 1.1 in)

Weights with cable

Option 105
Option 106

0.4 kg (0.88 Ib)
0.6 kg (1.32 Ib)

Fixed sensor cable
lengths

Option 105
Option 106

1.5 m (5-feet)
3.0 m (10-feet)

1. Beyond 70 % humidity, an additional 0.6% should be added to these values.

Environmental conditions

General

Complies with the requirements of the EMC Directive 89/336/EEC

Operating
Temperature

0° C to 55° C

Maximum humidity

95% at 40° C (non-condensing)

Minimum humidity

15% at 40 °C (non-condensing)

Maximum altitude

3,000 meters (9,840 feet)

Storage
Non-operating storage temperature –30° C to +70° C
Non-operating maximum humidity

90% at 65 °C (non-condescending)

Non-operating maximum altitude

15,420 meters (50,000 feet)