Specifications

Specifications describe the instrument's warranted performance. Non-warranted values are described
as typical. All specifications apply after a 30 minute warm-up phase with 50

W source/load resistance

and separate channels. All specifications are valid from 0°C to 55°C ambient temperature.

Timing Characteristics

Measured at 50% amplitude at fastest transitions in continuous mode and 50

W source impedance.

Mainframe

Output module

Agilent 8133A

Pulse Channel 1

Delay Pulse Channel 1

Option 001

Pulse Channel 2

Option 003

Pulse/Data Channel 2

Option 002

Frequency Range

33.0 MHz to 3.0000 GHz

Period Range

333 ps to 30.303 ns

Timing resolution

3.5 digits, 1 ps best case

RMS Jitter (period, delay, width)

less than 5 ps (1 ps typical)

10 ps ( f<2 GHz: 5ps)

Accuracy

±± 0.5% (±± 0.1% typical)

Width range

[1]

150 ps to (period – 150 ps) maximum 10.000 ns

Resolution

1ps

Accuracy

±± 100 ps (±± 30 ps typical)

Duty Cycle

[2]

0% to 100%

Resolution

0.1%, best case 1

Add. Variable delay

Pulse mode: no

Square mode: 0.000 ns

10.000 ns

Pulse mode: -5.000

to + 5.000ns

Square mode: -5.000

ns to 15.000 ns

[3]

Pulse mode: no

Square mode: 0.000 ns to 10.000 ns

Channel 1 to Channel 2 Delay

[4]

Pulse mode: -5.000 ns to + 5.000 ns
Square mode: -5.000 ns to 15.000 ns

Resolution

1 ps

Accuracy

[5]

±± 150 ps (±± 30 ps typical)

[6]

±± 50 ps

Phase

[7]

0°° to 3600°°

-3600°° to 3600°°

-3600°° to 3600°°

-3600°° to 3600°°

Resolution

0.1°°, best case 1 ps

Skew

[8]

±± 5 ns

Notes:

[1]

The width can only be varied in Pulse mode. In square mode the duty cycle is fixed at 50%

[2]

Width and duty cycle are mutually exclusive. Duty Cycle settings and limits are subject to the same specifications and settings as

Width.

[3]

Delay variation Channel 2 to Channel 1

[4]

The interchannel delay between Channel 1 and Channel 2 is the programmed delay of Channel

[5]

Any parameter variation

[6]

Only delay variation

[7]

Delay and Phase are mutually exclusive. Phase settings and limits are subject to the same specifications and settings as

[8]

Delay or Phase, plus Skew must be within the Delay Range

Repeatability:

is typ. four times better

than accuracy.

Internal Clock Generation

The internal clock can be set in
frequency or period mode

External Clock

The external input signal determines
the timebase of the instrument. The
external period or frequency is
measured and displayed, allowing the
correct setting of Duty Cycle and
Phase also in External Clock mode.

External Frequency Counter Period
Range
:

333 ps to 30.303 ns (300 ps to

500 ns typical)

Resolution:

1 ps

Accuracy:

± 0. 1 %