General specifications, Optional accessories – Atec Fluke-789 User Manual

Page 2

Advertising
background image

Measurement Function

Range and Resolution

Best Accuracy

(% of reading + LSD)

V dc

400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V

0.1 % + 1

V ac (true-rms)

400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V

0.7 % + 2

mA dc

30.000 mA

.05 % + 2

A dc

1.000 A (0.440 A continuous)

0.2 % + 2

A ac

1.000 A (0.440 A continuous)

1 % + 2

Resistance

400.0 Ohms, 4.000 k, 40.00 k, 400.0 k, 4.0 M, 40 M

0.2 % + 1

Frequency (0.5 Hz to 20 kHz)

199.99 Hz, 1999.9 Hz, 19.999 kHz

.005 % + 1

Diode Test

2.000 V (shows diode voltage drop)

2 % + 1

Continuity

Beeps for resistance < approx. 100 ohms

Fluke Corporation
PO Box 9090, Everett, WA USA 98206

Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands

For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com/

©2002 Fluke Corporation. All rights reserved. All
copyrights are the property of their respective owners.
Printed in U.S.A. 10/2002 1999242 D-ENG-N Rev C

Fluke. Keeping your world

up and running.

General specifications

Maximum voltage applied between
any jack and earth ground:
1000 V RMS
Storage temperature:

-40 °C to 60 °C

Operating temperature:

-20 °C to

55 °C
Temperature coefficient:

0.05 x (speci-

fied accuracy) per °C (for temperatures
< 18 °C or > 28 °C)
Relative humidity:

95 % up to 30 °C;

75 % up to 40 °C; 45 % up to 50 °C;
35 % up to 55 °C
Vibration:

Random, 2 g, 5-500 Hz

Shock:

1 meter drop test

Safety:

Designed in accordance with

EN61010, ANSI/ISA S82.01-1994 and
CAN/CSA C22.2 No. 1010..1-92
Over-voltage Category III
Size:

50 mm H x 100 mm W x 203 mm L

(1.97 in H x 3.94 in W x 8.00 in L)
Weight:

600 g (1.3 lbs)

Battery:

Four AA alkaline batteries

Battery life:

140 hours typical

(measurement), 10 hours typical
(sourcing 12 mA)
Warranty:

3 years

Output Function

Range and Resolution

Drive Capability

Accuracy

(% of span)

DC Current Output

0.000 to 20.000 mA or 4.000 to 20.000 mA

24 V compliance,

(selectable at power-up)

or,

.05 %

(Internal battery operation)

Over-range to 24.000 mA

1,200 Ohms,

@ 20 mA

DC Current Simulate

0.000 to 20.000 mA or 4.000 to 20.000 mA,

1000 Ohms,

(selectable at power-up)

@ 20 mA

.05 %

(Ext. 24 Volt loop supply)

Over-range to 24.000 mA

24 V Loop Supply

Minimum 24 V

250 Ohms

> 24 V

@ 20 mA

Current Adjustment Modes

Manual: Coarse, Fine, 25 % and 100 % step
Automatic: Slow Ramp, Fast Ramp, 25 % step

Temperature range of 18 °C to 28 °C, for one year after calibration

Optional accessories

80T-IR Infrared Temperature Probe
80TK Thermocouple Module
80T-150U Universal Temperature Probe
TL20 / TL22 / TL24 / TL26 / TL28 /
TL40A Test Lead Sets
AC20 / AC80 Clips
TP20 Industrial Test Probes
80i-400 AC Current Clamp*
i-410 AC/DC Current Clamp**
i-1010 AC/DC Current Clamp**
C25 / C100 Cases
80K-6 and 80K-40 High Voltage Probes
85RF High Frequency Probes
PV350 Pressure Vacuum Module
Fiber Optic Meter FOM
LPAK80-4 LockPak
TPAK ToolPak Meter Hanging Kit

*Above 50 A ac
**Above 1 A dc or 20 A ac

Ordering information

Model
Fluke-789

ProcessMeter
Including:
TL71 Premium Test
Lead Set plus
AC72 Alligator Clips,
Four AA alkaline
batteries (installed),
Users Manual and
Quick Reference
Guide