Preselection and preamplifier (r&s, Esrp-b2 option), Rf preamplifier (r&s – Atec Rohde-Schwarz-ESRP Series User Manual

Page 6: Fsv-b22 option), If and resolution bandwidths

Advertising
background image

Version 01.01, January 2013

6

Rohde & Schwarz R&S

®

ESRP EMI Test Receiver

Preselection and preamplifier (R&S

®

ESRP-B2 option)

Preselection
State

receiver mode

always on

analyzer mode

on/off (selectable)

Number of preselection filters

16

Bandwidths (–6 dB), nominal

10 Hz to 150 kHz

fixed lowpass filter

150 kHz to 30 MHz

35 MHz, fixed bandpass filter

30 MHz to 80 MHz

94 MHz, fixed bandpass filter

80 MHz to 130 MHz

94 MHz, fixed bandpass filter

130 MHz to 180 MHz

91 MHz, fixed bandpass filter

180 MHz to 230 MHz

105 MHz, fixed bandpass filter

230 MHz to 300 MHz

110 MHz, fixed bandpass filter

300 MHz to 425 MHz

195 MHz, fixed bandpass filter

425 MHz to 570 MHz

200 MHz, fixed bandpass filter

570 MHz to 715 MHz

210 MHz, fixed bandpass filter

715 MHz to 860 MHz

200 MHz, fixed bandpass filter

860 MHz to 1005 MHz

200 MHz, fixed bandpass filter

1005 MHz to 1750 MHz

fixed highpass filter

1750 MHz to 2850 MHz

fixed highpass filter

2850 MHz to 4850 MHz

fixed highpass filter

4850 MHz to 7000 MHz

fixed highpass filter

Preamplifier

switchable

Location

in the signal path between preselection
and 1st mixer

Frequency range

1 kHz to 7 GHz

Gain

20 dB (nom.)

RF preamplifier (R&S

®

FSV-B22 option)

Preamplifier
Availability

instruments without R&S

®

ESRP-B2 option receiver, analyzer

instruments with R&S

®

ESRP-B2 option

analyzer with preselection = off only

Frequency range

100 kHz to 7 GHz

Gain

20 dB (nom.)

IF and resolution bandwidths

IF and sweep filters
Resolution bandwidths (–3 dB)

receiver mode or analyzer mode,
span ≥ 10 Hz

10 Hz to 10 MHz in 1/2/3/5 sequence

analyzer mode, span = 0 Hz

20 MHz, 28 MHz, 40 MHz additionally

Bandwidth uncertainty

< 3 %

Shape factor 60 dB:3 dB

< 5

EMI bandwidths (–6 dB)

standard

200 Hz, 9 kHz, 120 kHz, 1 MHz

with R&S

®

ESRP-B29 option

10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz
additionally

Bandwidth uncertainty

< 3 %

Shape factor 60 dB:6 dB

< 4

FFT filters (analyzer mode)
Resolution bandwidths (–3 dB)

span ≥ 10 Hz

10 Hz to 300 kHz in 1/2/3/5 sequence

Bandwidth uncertainty

< 3 % (nom.)

Shape factor 60 dB:3 dB

< 5 (nom.)

Advertising