Atec Rohde-Schwarz-ESRP Series User Manual

Page 8

Advertising
background image

Version 01.01, January 2013

8

Rohde & Schwarz R&S

®

ESRP EMI Test Receiver

Second harmonic intercept (SHI)

RF attenuation = 0 dB, level = –10 dBm, preselection = off

5

,

with R&S

®

FSV-B22 option: RF preamplifier = off

100 MHz < f

in

≤ 3.5 GHz

45 dBm (nom.)

with R&S

®

ESRP-B2 option,

RF attenuation = 0 dB, level = –15 dBm, preselection = on

6

, preamplifier = off

100 MHz < f

in

≤ 3.5 GHz

50 dBm (nom.)

with R&S

®

ESRP-B2 option,

RF attenuation = 0 dB, level = –10 dBm, preselection = on

6

, preamplifier = on

100 MHz < f

in

≤ 3.5 GHz

35 dBm (nom.)

with R&S

®

FSV-B22 option,

preselection = off

5

, RF preamplifier = on, RF attenuation = 0 dB, level = –40 dBm

100 MHz < f

in

≤ 3.5 GHz

25 dBm (nom.)

Displayed average noise level
(analyzer mode)

RF attenuation = 0 dB, preselection = off/on

7

, preamplifier = off, termination = 50 Ω,

log. scaling, normalized to 1 Hz RBW, RBW = 1 kHz, VBW = 3 kHz, zero span, sweep
time = 50 ms, sample detector, trace average, sweep count = 20, mean marker
R&S

®

ESRP3, R&S

®

ESRP7

9 kHz ≤ f < 100 kHz

< –130 dBm,–140 dBm typ.

100 kHz ≤ f < 1 MHz

< –145 dBm,–150 dBm typ.

1 MHz ≤ f < 1 GHz

< –152 dBm,–155 dBm typ.

1 GHz ≤ f < 3.6 GHz

< –150 dBm,–151 dBm typ.

R&S

®

ESRP7

3.6 GHz ≤ f < 6 GHz

< –148 dBm,–151 dBm typ.

6 GHz ≤ f ≤ 7 GHz

< –146 dBm,–149 dBm typ.

with R&S

®

ESRP-B29 option, RF attenuation = 0 dB, preselection = off/on

7

,

preamplifier = off, termination = 50 Ω, log. scaling, normalized to 1 Hz RBW,
RBW = 10 Hz, VBW = 10 Hz, zero span, sweep time = 500 ms, sample detector,
trace average, sweep count = 20, mean marker
R&S

®

ESRP3, R&S

®

ESRP7

10 Hz

< –90 dBm,–100 dBm typ.

20 Hz

< –100 dBm,–110 dBm typ.

100 Hz

< –110 dBm,–120 dBm typ.

1 kHz

< –120 dBm,–130 dBm typ.

with R&S

®

ESRP-B2 option,

RF attenuation = 0 dB, preselection = on, preamplifier = on, termination = 50 Ω,
log. scaling, normalized to 1 Hz RBW, RBW = 1 kHz, VBW = 3 kHz, zero span,
sweep time = 50 ms, sample detector, trace average, sweep count = 20, mean marker
R&S

®

ESRP3, R&S

®

ESRP7

9 kHz ≤ f < 100 kHz

< –150 dBm,–155 dBm typ.

100 kHz ≤ f < 1 MHz

< –155 dBm,–160 dBm typ.

1 MHz ≤ f < 1 GHz

< –165 dBm,–168 dBm typ.

1 GHz ≤ f < 3.6 GHz

< –162 dBm,–165 dBm typ.

R&S

®

ESRP7

3.6 GHz ≤ f < 6 GHz

< –160 dBm,–163 dBm typ.

6 GHz ≤ f ≤ 7 GHz

< –158 dBm, typ. –161 dBm

with R&S

®

ESRP-B2 and R&S

®

ESRP-B29 option,

RF attenuation = 0 dB, preselection = on, preamplifier = on, termination = 50 Ω,
log. scaling, normalized to 10 Hz RBW, RBW = 10 Hz, VBW = 5 Hz, zero span,
sweep time = 500 ms, sample detector, trace average, sweep count = 20, mean marker
R&S

®

ESRP3, R&S

®

ESRP7

1 kHz

< –140 dBm,–150 dBm typ.

with R&S

®

FSV-B22 option,

RF attenuation = 0 dB, preselection = off

7

, RF preamplifier = on, termination = 50 Ω,

log. scaling, normalized to 1 Hz RBW, RBW = 1 kHz, VBW = 3 kHz, zero span,
sweep time = 50 ms, sample detector, trace average, sweep count = 20, mean marker
R&S

®

ESRP3, R&S

®

ESRP7

100 kHz ≤ f < 1 MHz

< –150 dBm,–155 dBm typ.

1 MHz ≤ f < 1 GHz

< –162 dBm,–165 dBm typ.

1 GHz ≤ f < 3.6 GHz

< –160 dBm,–163 dBm typ.

R&S

®

ESRP7

3.6 GHz ≤ f < 6 GHz

< –158 dBm,–161 dBm typ.

6 GHz ≤ f ≤ 7 GHz

< –156 dBm, typ. –159 dBm

7

With R&S

®

ESRP-B2 option. Preselection = off is only available in analyzer mode. In receiver mode preselection is permanently on.

Advertising