Atec Rohde-Schwarz-FSH3-6-18 User Manual

Specifications

Advertising
background image

R&S FSH3, R&S FSH6, R&S FSH18

Data Sheet

1145.5973.12

TI.1

E-14

Specifications

Specifications are valid under the following conditions: 15 minutes warm-up time at ambient temperature, specified environmental conditions met and
calibration cycle adhered to. Data without tolerances: typical values. Data designated as "nominal": design parameters, i.e. not tested.

Specification

Condition

R&S FSH3

R&S FSH6

R&S FSH18

Frequency

Frequency range

100 kHz to 3 GHz

100 kHz to 6 GHz

10 MHz to 18 GHz

Reference frequency

Aging

1 ppm/year

Temperature drift

0 °C to 30 °C
30 °C to 50 °C

2 ppm

in addition 2 ppm/10°C

Frequency counter

Resolution

1 Hz

Frequency span

0 Hz, 100 Hz to 3 GHz

0 Hz, 100 Hz to 6 GHz

0 Hz, 100 Hz to 18 GHz

1145.5850.13

0 Hz, 1 kHz to 3 GHz

-

-

Spectral purity
SSB phase noise

f = 500 MHz, 20 to 30 °C

30 kHz from carrier

<-85 dBc/(1 Hz)

<-85 dBc/(1 Hz)

100 kHz from carrier

< -100 dBc/(1 Hz)

< -90 dBc/(1 Hz)

1 MHz from carrier

< -120 dBc/1 Hz)

< -100 dBc/(1 Hz)

Sweep time

span = 0 Hz

1 ms to 100 s

span > 0 Hz

20 ms to 1000 s, min. 20 ms/600 MHz

Bandwidths

Resolution bandwidths
(-3 dB)

1145.5850.13

1, 3, 10, 30,100, 200, 300 kHz, 1 MHz

1145.5850.03, .23,
1145.5850.06, .26, .18

In addition 100, 300 Hz

Tolerance

300 kHz

± 5 %, nominal

1 MHz

± 10 %, nominal

Advertising