R&S FSH3, R&S FSH6, R&S FSH18
Data Sheet
1145.5973.12
TI.7
E-14
Accessories
Power Sensors R&S FSH-Z1 and R&S FSH-Z18
Frequency range R&S FSH-Z1
10 MHz to 8 GHz
R&S FSH-Z18
10 MHz to 18 GHz
VSWR
10 MHz to 30 MHz
30 MHz to 2.4 GHz
2.4 GHz to 8 GHz
8 GHz to 18 GHz
< 1.15
< 1.13
< 1.20
<1.25
Maximum input power
average power
peak power
(<10 Is, 1% duty cycle)
400 mW (+26 dBm)
1 W (+30 dBm)
Measurement range
200 pW to 200 mW (-67 dBm to +23 dBm)
Signal weighting
average power
Effect of harmonics
Effect of modulation
<0.5 % (0.02 dB) at harmonic ratio of 20 dB
<1.5 % (0.07 dB) for continuous digital modulation
Absolute measurement
uncertainty
sine signals,
no zero offset
10 MHz to 8 GHz
8 GHz to 18 GHz
15 °C to 35 °C
0 °C to 50 °C
15 °C to 35 °C
0 °C to 50 °C
<2.3 % (0.10 dB)
<4.2 % (0.18 dB)
<3.5 % (0.15 dB)
<5.0 % (0.21 dB)
Zero offset after zeroing
< 110 pW
Dimensions
48 mm x 31 mm x 170 mm, connecting cable 1.5 m
Weight
< 0.3 kg