2
ITS 6006 datasheet
2
Radiated immunity test system 6 GHz
its 6006
Technical specifications
Generator
RF
Frequency range:
80 MHz – 6 GHz
Resolution:
1 Hz
Reference frequency:
10 MHz
Max. adjustment deviation: ±50 Hz
Aging:
5 ppm
Level range:
-60 dBm to +10 dBm (-60 dBm to +8 dBm at f>3 GHz & ϑ>30°C)
Resolution:
0.1 dB
Accuracy:
<1 dB, typ. <0.5 dB (+15 to +25°C)
Spectral distortion
Harmonics (level ≤0 dBm):
<-30 dBc, typ. <-35 dBc
Harmonics (level ≤+10 dBm): <-25 dBc
Non harmonics:
<-35 dBc
Amplitude modulation
Modulation depth:
0 – 100%
Resolution:
0.5%
Modulation frequency range: 1 Hz – 20 kHz
Frequency resolution:
1 Hz
ITS 6006 Block diagram
Signal generator
4 RF outputs
4x Interlock
User-Port-IN 0-3
Digital 1 & 2
Optical input
Analog input
External
modulation
10 MHz
reference
Trigger
User-Port-OUT 0-3
RF switch
2x SPDT
3x PMR 6006
power meter
Remote control:
LAN, RS232, USB
Control unit
EUT monitoring interface