91

Section 5

Performance Check Procedures

5.1 Introduction

This section provides the procedure for checking the electrical performance requirements of the Model

4076 and 4079 Arbitrary Waveform Generators, as listed in instrument Operating Manual Section 1: “

Specifications”.
If the waveform generator fails to meet these checks, then you should perform the adjustment

procedure. Adjustment procedures are available on a separate document.
The Performance Check Procedures are identical for both channels on the Model 4079 and must be

performed separately for each channel.
At the end of this document, a table is provided to record the test results.

5.2 Test Equipment


The following table lists the equipment necessary to perform the performance tests. Any equivalent

equipment may be substituted for the recommended model.

Description

Minimum Requirements

Recommended Model


Oscilloscope

Bandwidth > 400MHz

Tektronix DPO4054

Distortion Analyzer

Frequency range 10Hz-100KHz Tektronix AA5001

Digital Voltmeter

6 1/2 digits, true RMS

Fluke 8845A

Universal Counter

200MHz, Time A B, Period

Pendulum CNT-90

Function Generator

10MHz output

BK Model 4045

50Ω Feed-through

±1% to 1GHz (high frequency)

Tektronix 011-0049-02

Precision 50Ω Feed-through

±0.1% (DC to 100KHz)

Tektronix 012-0482-XX

50Ω Coaxial Cables, with BNC connectors, 1 m,