Set to factory default, 5 set to factory default – B&K Precision 2516 - Manual User Manual

Page 32

Advertising
background image

32


To perform self-calibration, follow these steps:

1.

Power ON the instrument and allow it to warm up for 30

minutes.

2.

Press

and press

to select Do Self Cal.

3.

Follow the prompt on the screen and press F4 to proceed. To

exit and cancel self-calibration, press F5.

4.

The following screen will display. Wait for the progress bar to

show 100% complete. When self-calibration is finished, the
screen will prompt you to exit.


2.5 Set to Factory Default

To set the instrument to factory default settings, press

and toggle

to select Type and Factory. A Load option will appear. Press the

corresponding

key to load factory default settings.

Table 1 - Factory Default Settings