B&K Precision 2516 - Manual User Manual

Page 52

Advertising
background image

52

Window

(see table

below)

Selects FFT windowing.

FFT Zoom

1X, 2X, 5X, 10X Zoom function of FFT waveform.


Table 11 - FFT Window Options

Window

Characteristic

Best Use Cases


Rectangular

The best frequency
resolution but the
worst magnitude
resolution. It is
essentially the same as
having no window.

Symmetric transients or bursts.
Equal-amplitude sine waves
with fixed frequencies.
Broadband random noise with
a relatively slowly varying
spectrum.

Hanning

Better frequency,
poorer magnitude
accuracy than
rectangular
window

Sine, periodic, and narrow-
band random noise.

Hamming

Hamming has a slightly
better frequency
resolution than
Hanning.

Transients or bursts pulse. The
ranges of the signals have great
difference from before to after.

Blackman

Best magnitude
resolution but worst
frequency resolution.

Single frequency waveforms, to
find higher order harmonics.

Table 12 - FFT Menu 2/2

Menu Option

Setting

Description

Scale

dBVrms

Set unit scale to dBVrms

Vrms

Set unit scale to Vrms

Display

CH1

Selects channel 1 as FFT source.

CH2

Selects channel 2 as FFT source.

ToZero

Selects FFT windowing.

Advertising