2-3

Installation and Operation Manual
X-VA-1110-1140-eng
Part Number: 541B040AAG
September, 2012

Brooks

®

1110 and 1140 Series

Section 2 Installation

Figure 2-1 Typical Bypass Installation

A - Inlet Valve

B - Outlet Valve

C - Bypass Valve

D - Control Valve

E - Drain Valve

HORIZONTAL

LINE

D

A

E

C

VERTICAL

LINE

B

B

A

E

FLOWMETER

C

D

FLOWMETER

A. Location
For proper operation of the Full-View flowmeter it must be mounted within
6 degrees of true vertical, with the inlet connection at the bottom of the
meter, and the outlet at the top. The use of a level is recommended to
assure vertical positioning.

B. Piping Arrangement
Installation of a bypass piping arrangement is strongly recommended,
Figure 2-1. Bypass piping permits the meter to be isolated from the flow
line for servicing and cleaning. For more details refer to Brooks T-023
Technical Bulletin Guide for bypass meters.

2-7 Installation of meter