Warning, Brooks, Glass tube explosion hazard – Brooks Instrument 1100 Series User Manual

Page 9

Advertising
background image

1-3

Installation and Operation Manual
X-VA-1110-1140-eng
Part Number: 541B040AAG
September, 2012

Brooks

®

1110 and 1140 Series

Section 1 Introduction

Table 1-1 Pressure Ratings and PED Category

GLASS TUBE

EXPLOSION

HAZARD

WARNING

Plastic protective sleeve must remain over glass
tube. (Meter sizes 7 -13 only)
Fasten meter windows securely.
Do not operate above pressure and temperature
limits.
Avoid pressure and flow surges.
Do not service or repair while pressurized.
Read and understand instruction manual.
Failure to comply could result in serious personal
injury or property damage.

End

Maximum Operating Pressure (psig)

2

Meter

Fitting

@ 70°F

@ 150°F

1

@ 250°F

PED

Size

Material

(21°C)

(66°C)

(121°C)

CAT.

2

CPVC

100

100

Not Available

Metal

500

500

500

316 SS

(w/CRN)

200

200

200

6

CPVC

100

100

Not Available

Metal

450

450

450

316 SS

(w/CRN)

200

200

200

7

CPVC

100

100

Not Available

Metal

350

350

350

SEP

8

CPVC

100

100

Not Available

Metal

300

300

300

316 SS

(w/CRN)

160

160

160

9

PVC

100

100 (Note 1) Not Available

Metal

175

175

175

316 SS

(w/CRN)

100

100

100

10

PVC

100

100 (Note 1) Not Available

Metal

100

100

100

316 SS

(w/CRN)

60

60

60

12

Metal

75

75

75

SEE

NOTE

13

Metal

100

100

100

3

NOTES:

1

Maximum temperature rating for nonmetal end fittings is 150°F

(66°C) for all materials except PVC, which is 140°F.

2

Flanged unit pressure ratings are limited to ANSI flange or tube ratings

whichever is lower.

3

Sizes 12 and 13 do not conform to Pressure Equipment Directive

97/23/EC, therefore it cannot be sold or used in the EU/EFTA.

Advertising