Installation of sensor

Fluid Components Intl

Document 06EN003327 Rev. -

23

FlexCOR

Model CMF Series

4.5

Vibrations

Locate the flowmeter as far away as possible from components that generate mechanical vibration

in the piping.

Or ensure that there is no direct connection with

them e.g. by using flexible connections.

The flowmeter can also be located after a bend.

4.6

Cross-talk

If the flowmeters are located close to each other, e.g. in the same pipe section, the meters may

disturb each other in measurement, especially with low flow. Locate the meters with a flexible
connection instead of a permanent connection.

Avoid mounting the meter on the same steel frame. i.e. insulate the meters mechanically.

4.7

Magnetic fields

Locate the sensor a minimum of 10 inches (25 cm) from strong magnetic fields (motors,
transformers, electrically operated valves, etc.).

4.8

Transportation/

storage

The sensor is a fragile piece of equipment and shall be placed in its storage carton when

transported or stored. If this is not possible, the sensor must be packed so the packing enclosure
can withstand the hazards from transportation or storage.