Fluid Components International CMF Series Manual Installation of Flow Element User Manual

Page 5

Advertising
background image

Installation of sensor

FlexCOR

Model CMF Series

26

Fluid Components Intl

Document 06EN003327 Rev. -

4.12

Location

To ensure the optimum function of the measuring equipment it is important that the

installation instructions are followed closely.

The flowmeter can be located both indoors and

outdoors, but the following conditions must be
observed:

Liquid temperature

The CMF - A is available in 2 versions.

Standard version: -40 to 257 °F (-40 to +125°C).
High temperature version: -40 to 356°F (-40 to +180°C).

For the high temperature version the multiple plug is raised from the sensor housing by a pipe.
It is possible to insulate the sensor and still having access to the plug.

Important

When the temperature difference between a liquid and the surroundings is large, the sensor must
be insulated to prevent 2-phase flow and the loss of measuring accuracy. This applies especially

in the case of low flow.

The sensor must always

be completely filled with a homogeneous liquid or gas in single phase,

otherwise measuring errors will occur.

If there is air/gas in the liquid or liquids which are volatile, horizontal sensor mounting is

recommended.

4.13

Mounting

The mounting bracket supplied with the unit must always be used. The bracket must be mounted

on a wall or steel frame which is vibration free and mechanically stable.