Fluid Components International CMF Series Manual Wiring Diagrams User Manual

Page 2

Advertising
background image

Elec

trcical

con.

Fluid Components Intl

Document 06EN003327 Rev. -

FlexCOR

Model CMF Series

46

6.2

Transmitter with

extended output’s

(only 19” version),
terminal board

083H4253 &

083H4255

Installation

1) Mains supply 115 to 230 V a.c. from building installation Class II. A switch or circuit-breaker

(max. 15 A) shall be included in the building installation. It must be in close proximity to the

equipment and within easy reach of the OPERATOR, and it shall be marked as the

disconnecting device for the equipment.

2) Protective conductor terminal. Required cable min. AGW16 or 1.5 Cu.

The insulation between the connected mains supply and 24 V a.c./d.c. supply for the flow-
meters, models 24 V a.c./d.c. shall at least be rated with double or reinforced insulation at

mains voltage.
For field wiring installation National Installation Code shall be met of the country, where the

flowmeters are installed.

Digital output

If the internal resistance of the loads exceeds 10K ohms, connect an external 10K ohm load
resistor in parallel to the load.