Elec

trcical

con.

Fluid Components Intl

Document 06EN003327 Rev. -

FlexCOR

Model CMF Series

50

6.8

Setting of active or

passive current out-

put mode

The current output of the transmitter can

operate in either active or passive mode to
make elec-trical connection as easy as

possible. The default current output in the

transmitter is set to passive mode and must
be looped powered.

If an active mode is required, a jumper on the

transmitter PCB must be put in active position.
This is done by taking out the transmitter elec-

tronics, follow the instructions in section

5.2.11 or 5.2.12. Jumper position is shown
below.

Passive mode: Jumper in right position.
Active mode: Jumper in left position.

Be aware that in active mode the output shall be considered as a barrier output. The connection

is not safe when put in active mode and accidentally connected to a barrier intended for use in
passive mode.

Important

6.9

Installation

examples

Current output in passive

mode

Frequency/pulse output in

passive mode

HART output