Section 4 system startup, 1 initializing the instrument – Hach-Lange PHOSPHAX indoor sc User Manual

Page 33

Advertising
background image

33

Section 4

System Startup

4.1 Initializing the instrument

Important Note: The instrument only can work correctly if it is at
operating temperature. Let the instrument warm up for a minimum
of an hour so the inside of the enclosure, chemicals and electrode
are at operating temperature.

1. Make sure the PHOSPHAX sc is registered in the sc1000

system. If necessary, initiate the controller to search for the
analyzer. Refer to the sc1000 user manual for more
information.

2. Configure the analyzer in the SENSOR SETUP menu and note

the settings. Refer to

section 5.2 on page 35

for more

information. The factory settings (default setting) are
appropriate for most typical applications.

3. From SENSOR SETUP, select

PHOSPHAX sc>MAINTENANCE>TEST/MAIN.

4. Select the PREPUMP ALL function and confirm.

5. Wait until the analyzer returns to the service state (displayed in

TEST/MAIN.>PROCESS) after the prepumping sequence is
completed.

6. Select START from the maintenance menu.

Advertising
This manual is related to the following products: