Hach-Lange SI792 C Quick start guide User Manual

Quick start guide

Advertising
background image

© Hach Company, 2007. All rights reserved. Printed in Germany.

Edition 1 December 2007

Quick start guide

si792 C 2-wire conductivity transmitter

DOC016.53.00795

4–20 mA signal values

Default settings:

• 0 mS at 4 mA
• 100 mS at 20 mA

1. Press

CONF

, enter

1200

for the “out. 1MNU” menu.

2. Press

ENTER

to go to the "o1.4mA" or "o1.20mA" sub-menu.

3. Set the values for 4 mA and 20 mA.
4. Press

CONF

, then

ENTER

to exit the configuration mode.

Enter the cell constant

1. Press

CAL

and enter passcode: 1100,

ENTER

. CAL CELL will be displayed for 3 seconds.

2. Use the

UP ARROW

and

RIGHT ARROW

to enter the value of the cell constant of the con-

nected sensor. The conductivity value will be displayed also.

Note: For Hach or GLI contacting conductivity sensors, the cell constant is usually printed on a
tag with the sensor wiring.
Note: If the entry of the cell constant takes longer than 6 seconds, the display will alternate
between showing the conductivity and the temperature value.

3. Press

ENTER

to save the value.

4. The transmitter displays the cell constant value and CELL. Press

ENTER

.

5. The transmitter remains in HoLD mode. Press

ENTER

again.

The transmitter remains in HoLD mode for approximately 20 seconds after calibration and
returns to the measure mode.

Note: Repeat entry of the cell constant when “Err” is displayed.

Calibration

The sensor can be calibrated with a calibration solution or by comparison.

Calibration with a calibration solution

Use a calibration solution with a known conductivity value to adjust the transmitter to show
the same value as the solution.

Note: Stabilize the temperature during calibration.

1. Press

CAL

, enter passcode: 0110,

ENTER

. CAL SOL will be displayed for 3 seconds. The

transmitter is ready for calibration.

2. Remove and clean the sensor.
3. Immerse the sensor in the calibration solution.
4. Use the

UP ARROW

and

RIGHT ARROW

to enter the known value of the calibration solu-

tion.

Note: If the entry of the calibration solution value takes longer than 6 seconds, the display will
alternate between showing the cell constant and the temperature value.

5. Press

ENTER

to save the value.

6. The display will show the new cell constant. Press

ENTER

.

7. The transmitter remains in HoLD mode. Press

ENTER

again.

The transmitter remains in HoLD mode for approximately 20 seconds after calibration and
returns to the measurement mode.

Note: Repeat the calibration when “Err” is displayed.

DANGER
Explosion hazard. This Quick Start Guide is intended for setup of the si792
transmitter in a non-hazardous location. It is not a substitute for the instructions in
the user manual (DOC026.53.00795). Installation in a hazardous location must follow
the applicable control drawing or ATEX certificate (refer to the documentation CD).

DANGER: Electric shock hazard. Disconnect power to all wiring before making
connection to the si792 C series transmitters.

Connect 24 VDC power and communications

Remove the power source before making any electrical connections to the si792 transmitter.

Connect the sensor

For sensors other than the Hach 3400 series or 831x, refer to the user manual.

Basic configuration—4–20 mA

1. Press

CONF

and enter

1200

to enter the configuration mode. The following menus can

be changed:

• 4–20 mA current output—out.1MNU
• Temperature compensation—tc MNU
• Alarms—ALrt MNU

2. To move to a different menu, press the

RIGHT ARROW

">", then press enter.

3. To change a setting within a menu, use the

UP ARROW

"^" or

RIGHT ARROW

">" buttons

to change the setting.

4. Press

ENTER

to move to the next item in the menu.

To exit the configuration mode at any time, press

CONF

and then

ENTER

. Refer to the user

manual for additional configuration options.

Terminal No.

si792 transmitter version

4–20 mA/HART

Profibus PA

Foundation Fieldbus

10

(–)

(–)

(–)

11

(+)

(+)

(+)

Jumper

Terminal No.

3400 series color

831x series color

Jumper between 1 and 2

1

2

black

white/yellow

Jumper between 3 and 4

Jumper between 4 and 5

3

red

red

4

5

clear

white/orange

7

blue

blue

8

white

black