Calibration by comparison/grab sample

The sensor can be calibrated by measuring the process water with a portable instrument or
taking a grab sample and measuring the value in the lab (one-point calibration). The sensor
does not need to be removed from the process during calibration. The sample temperature
should correspond to the measured process temperature for accurate results.

Note: The calibration by comparison option is not available when the range and units are set
to 00.00%.

1. Press

CAL

, enter passcode: 1105,

ENTER

. The display will show CAL PRD and then StorE.

2. Measure the process water with a portable instrument or collect a grab sample for mea-

surement in the lab. Press

ENTER

to proceed.

The cal mode indicator will flash to indicate that calibration has not been completed. The
transmitter will continue to measure and display the current sample values.

3. Measure the grab sample with a laboratory or portable meter.
4. Press

CAL

, enter passcode: 1105,

ENTER

to access the product calibration once more.

The display will briefly show CAL PRD and then CALC.

5. Edit the displayed value to match the value measured by the laboratory or portable meter

and press

ENTER

. The display will show the new cell constant.

Press

ENTER

to end the calibration. The display will alternate between the measured value

and Hold. Press

ENTER

to return to the measuring mode. The outputs will remain in the hold

mode for approximately 20 seconds.

Note: For comparison calibration via Foundation Fieldbus, see the user manual.

Table 1 Default passcode settings

Display

Function

Default setting

Cal INFO

Calibration information

0000

CAL CELL

Calibration mode for si792x C-PA

1100

CAL SOL

Calibration with standard solution

0110

1

CAL PROD

Product calibration

1105

1

CAL RTD

Temperature sensor adjustment

1015

CFG ERR

Error information

0000

CFG CONF

Configuration mode

1200

CFG SNSR MO

Sensor monitor

2222

CFG OUT

Current source

5555

CFG SPCL ST

Administrator passcode. Use

ARROW

keys to

select NO or YES

1989

NO SPCL ST

Do not change
administrator
passcode

Press

ENTER

to return to the default

Press

CONF

to exit the editor

YES SPCL ST

Change administrator
passcode

Press

ENTER

to accept the new

Press

CONF

to exit the editor

1

Not available with si792x C-PA