Operation—Foundation Fieldbus

43

Select temperature compensation

1

tc.

OFF

Temperature compensation turned off

LIN (default)

Linear temperature compensation
Enter temperature coefficient:

Range: 0–19.99%/K (default: 2.00%/K)
Reference temperature = 25 °C

nLF

Natural waters (to EN 27888)

nACL

Ultrapure water with NaCl traces (0–120 °C)

HCL

Ultrapure water with HCl traces (0–120 °C)

nH3

Ultrapure water with NH

3

traces (0–120 °C)

Select Sensocheck

AL.SnSO

CHECK ON

Continuous Sensocheck evaluation of
sensor function

CHECK OFF (default)

LED in Hold mode

AL.LED

HOLD ON

LED blinks during hold

HOLD OFF (default)

LED off during hold

Enter Fieldbus address (optional)

2

FF.ADR

0026 BUS (default)

Range: 0017–0031

1

Temperature compensation is not available for salinity, concentration or USP.

2

Use only when there is no bus connection. The transmitter will restart and set all

parameters to default values. Individual settings must be entered once more.

6.1.2

Configuration menu (continued)