Operation—Foundation Fieldbus

48

6.2.2.3 Data processing

Use the L_TYPE parameter to apply a linearization function to the data.

Direct—data is sent directly from the TB to the AI without
processing. The units for the XD_SCALE and OUT_SCALE
parameters must be identical.

Indirect—data from the TB is linearly scaled from the input scale
(XD_SCALE) to the output scale (OUT_SCALE).

Indirect square root—data is rescaled from the input scale
(XD_SCALE) and recalculated using a root function. Then the
value is linearly scaled to the output scale (OUT_SCALE).

6.2.2.4 Alarms

The AI block can generate block alarms and limit alarms. Use the
ACK_OPTION parameter to specify if an alarm must be acknowledged.
When the measured value status is “bad”, the AI block BLOCK_ERR
parameter indicates an Input Failure.

Block alarms—a block error will be reported via the
BLOCK_ERR parameter (simulate active, input failure, block
configuration error, out of service (OOS)). The BLOCK_ALM
parameter sends the alarm status to the control system.

Limit alarms—the measured value OUT falls outside of the limit
values (HI_HI_LIM, HI_LIM, LO_LIM, LO_LO_LIM).

If an alarm occurs, evaluate the following bus parameters:

OUT parameter (currently measured value) in the Al block

LAST_ERROR parameter in the transducer block

SENSOFACE_STATUS parameter in the transducer block