4 modbus card connections, Figure 23, Installation – Hach-Lange SC 1000 User Manual

Page 35

Advertising
background image

33

Installation

5. Write the serial number from the rating plate on the supplied sticker and attach it to

the main high voltage barrier (

Figure 18

).

6. Install the probe module cover.

After installation and connection of a plug-in expansion card, the card must be configured
to the system. For output card setup instructions, refer to

section 6.3.1, page 78

.

3.6.4 Modbus card connections

Modbus RS485 (YAB021) and Modbus RS232 (YAB047) are available. For more detailed
information refer to the bus system manual.

To make a Modbus card connection:

1. Remove power from the instrument. Remove the probe module cover.

2. Connect the Modbus card to the appropriate slot (

Figure 18

). Use a magnetic

screwdriver to secure the four screws to the card.

3. Install the card connector to the appropriate connection on the main circuit board

(

Figure 17

).

4. Feed the cable through the base of the module and properly prepare and insert each

wire into the terminal according to

Figure 24

/

Table 7

and

Figure 25

/

Table 8

.

5. Write the serial number from the rating plate on the supplied sticker and attach it to the

main high voltage barrier (

Figure 18

).

6. Install the probe module cover.

Figure 23 Output card (YAB019) cable connections

1

Terminal Block

Refer to

Table 6

for terminal assignments.

Table 6 Output card (YAB019) terminal assignments

Terminal

Designation

1

Output 1+

2

Output 1 –

3

Output 2 +

4

Output 2 –

5

Output 3 +

6

Output 3 –

7

Output 4 +

8

Output 4 –

9

Shield (Connected to protective earth)