Advanced operations – Hach-Lange SC 1000 User Manual

Page 99

Advertising
background image

97

Advanced Operations

Figure 61

shows the behavior of the relay in the Warning mode under different conditions.

PROCESS EVENT

Default value: Disabled
Sets the monitoring of the internal process event bits of the selected source.
ENABLED: Monitoring is active.
DISABLED: Monitoring is not active.

SET TRANSFER

Default value: DE-ENERGIZED
Sets the status of the relay (ENERGIZED/DE-ENERGIZED) if some or all conditions (this means
warning, error or process event bits) are detected in the selected source or if the source is
missing.

ON DELAY
(0 s

999 s)

Default value: 5 seconds
Sets a delay time for the relay to turn on.

OFF DELAY
(0 s

999 s)

Default value: 5 seconds
Sets a delay time for the relay to turn off.

WARNING

Figure 61 Relay behavior—Warning mode (provided Error List and Warning List are enabled)

1

Bit set

3

Source (y-axis)

2

Time (x-axis)