Zero calibration – Hach-Lange SI794 D User Manual User Manual

Page 70

Advertising
background image

Zero calibration

Zero calibration

The sensors have a very low zero current. Therefore, a zero cali-

bration is only recommended for measurement of oxygen traces.

If a zero calibration is performed, the DO sensor should remain

for at least 10 to 30 minutes in the calibration medium in order to

obtain stable, non-drifting values.

During zero calibration, a drift check is not performed.

Zero current of a properly functioning sensor is notably less than

0.5 % of air current. The display (bottom: measured value, top:

entered value) does not change until an input current is entered

for the zero point.

When measuring in an oxygen-free medium, the displayed

current can be taken directly.

70