Grab sample, Maintenance/‍diag, Reagent changes – Hach-Lange POLYMETRON 9245 Basic User Manual User Manual

Page 21: See section entitled, Maintenance/diag

Advertising
background image

6. Place the lid back on top of the overflow vessel and select OK to start the measurement. The

display switches back to the main measurement screen and displays the measurement value.

7. Once measurement of the solution is complete, the measured value and time will be displayed

against verif. s. This should be almost identical to the value entered in the VALID. SOL field
previously.
The value displayed in gap, will be the percentage difference between the concentration of the
verification solution entered in the VALID. SOL field, and the concentration measured by the
instrument. If this value is too high, then a calibration of the analyzer is recommended.

8. Once the cycle is complete, the analyzer automatically reverts back to standard measurement

mode.

9. Select Main to go back to the main measurement screen.

GRAB SAMPLE

Similar to the verification process, this option allows measurement of a sample of your choice. Before
starting this process, ensure you have a sample of approximately 200 mL to measure.

Note: It is recommended that the grab sample should have a concentration of > 10 ppb and should be at the same
temperature as during calibration for better accuracy. Under these conditions, measurement with manual
introduction gives 5% accuracy from 10 ppb to 10 ppm at a temperature which is within ± 5 °C of the temperature
during calibration.

1. The first step of this cycle allows the reactivation of the electrode. The message SENSOR

ACTIVATION is displayed. Select Yes to reactivate the sensor. Select No if time is more
important than accuracy.

2. The reactivation process is automatically followed by a rinsing of the overflow vessel and

measuring cell using the process sample. Upon completion of the rinsing step, the analyzer is
ready for the first manual step in the cycle.

3. Place the lid of the overflow vessel on the side and pour approximately 200 mL of the solution

into the overflow vessel, as prompted on screen. Manual introduction is complete when the
sample overflows at the back of the overflow vessel.

4. Place the lid back on top of the overflow vessel and select OK to start the measurement. The

display switches back to the main measurement screen and displays the measurement value.

5. Once measurement of the solution is complete, the measured value and time will be displayed

against grab s.

6. Once the cycle is complete, the analyzer automatically reverts back to standard measurement

mode.

7. Select Main to go back to the main measurement screen.

MAINTENANCE/DIAG.

Reagent changes

This procedure is applicable to each change of reagent solution. It allows the regulation of the
volumes of solutions in the bottles. The analyzer consistently checks the consumption of these
solutions, triggering an alarm when these levels get too low. This option must be selected each time
a solution is refilled or changed to reset the measurement parameters.

Note: The conditioning reagent and electrolyte consumption are approximate measurements, so a discrepancy
between measurement and actual consumption may occur. The conditioning solution consumption has been
measured for a pH of 11.2 at an ambient temperature of 25°C.

English 21

Advertising