Change the language, Standard operation, Calibration – Hach-Lange SENSION+ MM340 User Manual

Page 9: Calibration settings, Calibration types

Advertising
background image

Change the language

The display language is selected when the meter is powered on for the
first time.

1. Use the or to select a language from a list.
2. Confirm with . The measurement screen shows DATA OUTPUT.
3. Select Deactivated if no printer or PC is connected and confirm.

Refer to

Select the data output

on page 11 for more information

about Data Output.

Standard operation

Calibration

W A R N I N G

Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.

Calibration settings

The calibration settings contain Calibration type, Calibration frequency
and Display options.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select the following options:

Option

Description

Stability C.:

Criterion by stability—select Fast, Standard or Strict.

Calibration type

Calibration type—select Technical buffers,
DIN19266 Buffers, User Buffers, To a X value, data
introduction or Theoretical calibration. Refer to

Calibration types

for more information.

Option

Description

Cal. frequency

Calibration reminder—can be set between 0–7 days
(default daily). The display shows the remaining time to
the new calibration. Refer to

Set the calibration reminder

on page 10 for more information.

Display mV

Display mV—select YES or NO to show mV.

Standard 220 mV Standard 220 mV—calibrate with a 220 mV standard

solution

To a X value

To a X value—use a specific standard solution. Adjust
the value during the calibration.

Data introduction Data introduction Compensation—enter the

compensation value

Factory adjust

Factory adjust—enter the compensation value

Calibration types
Different calibration types can be selected.

1. From the main menu use the or to select CALIBRATION.

Confirm.

2. Use the to enter the calibration menu.
3. Use the or to select Calibration type.

Option

Description

Technical buffers

pH 2.00, 4.01, 7.00, 9.21 and 10.01 at 25 °C
(77 °F)

DIN19266 Buffers

pH 1.679, 4.006, 6.865, 7.000, 9.180, 10.012 and
12.454

User Buffers

Selected when the technical or DIN19266 buffers
are not used. Refer to

Technical buffer solutions

(DIN 19267)

on page 18 for pH values of

specific buffer sets at varying temperatures.

Calibration to a X value To adjust manually any scale value of the

measured pH.

English 9