RANGE JUMPER SELECTION
DIGIT SELECT KEY
VALUE SELECT KEY
Press
to select short.
Press the
key to display
InPut (Input type selection).
Press the
key until rESEt is displayed. Short the input leads at the RTD sensor. Press
the
key to store value. Set up for RTD per section 14.3
Press
to display input
selected. Press
again
until rtd is displayed.
14.2
2-WIRE RTD LEAD COMPENSATION
This section describes how to remove the error caused by lead resistance in a 2-wire RTD.
Ambient temperature changes will cause some error in the readings; the higher the lead
resistance, the greater the error. When performing this procedure, the leads should be shorted
together as close as possible to the RTD. This step is not necessary when using 3- or 4-wire
RTD's since lead resistance compensation is automatic in the meter. When setting up the
meter, it may be necessary to enable some of the menu items. See Section 9 for further
information.
MENU KEY
All RTD Types
Jumpers Required
2-,4-wire
b, e
Figure 14.1 Temperature Signal Conditioner
Note: See Section 22 to select 10 Vdc excitation.
Jumpers Required
3--wire
a, e
a
e
f
b