Calibration ‘0-1’ function, Formula 2 – Liquid Controls IT275N User Manual

Page 7

Advertising
background image

Sponsler, Inc.

IT275N Remote Totalizer & Rate Indicator

pg. 7

DOC#: MN-275N-A


On the IT275F PCB

Set S5@4, S6, 7, 8@0 (.4000)

Set S4 in ‘0’ position (0.4000)


The electrical accuracy can be verified by injecting a stable frequency @ TB2-1,2 on the Mounting PCB
and incorporating the following formula:

Total = F x T x C.F.

D


Where F = Frequency in Hz (Frequency = K-Factor X Flowrate)

60


T = Time (Duration) of test in seconds

C.F. = Calibration Factor as entered in S4-S8

D = Divider as entered in S3


Example 2:

F = 500 Hz T = 1 minute (60 sec) C.F. = .4000

Total Displayed =

500 x 60 x .4000

=

30,000 x .400

=

12000

= 120 in 1 minute

100

100

100


Rate Indicator will display 120 continuously.


FIGURE 3















Calibration ‘0-1’ Function

The ‘0-1’ function provides enhanced accuracy when totalization encompasses a large quantity for an
extended period of time such as SCF produced in a 24-hour period.

The ‘0-1’ function should be incorporated only when both conditions listed below are met:
1) C.F.’s 1

ST

digit right of decimal is 1

2) C.F.’s 5

th

digit right of decimal is not 0

Formula 2