www.mastercool.com

9. Click on “GET DATA FROM MASTERCOOL WEBSITE”. Once

the data is downloaded, the “Download Data to Device”

square will turn green.
10. Click on “DOWNLOAD DATA TO DEVICE”. When the download

is done, (it only takes a fraction of a second) a window will pop

up telling you if the download was successful.
NOTE: If the refrigerant is already in the 52246, a pop up window

will tell you and the refrigerant data will not be downloaded.
11. After the refrigerant is downloaded, the 52246 will turn off.
If you would like to download another refrigerant, turn the 52246

back on and start from step 6 above.
If you are done loading refrigerants, reassemble the 52246, taking

care that the battery wires don’t get trapped between the case

halves.
After turning the unit on, press the down arrow to go to the last

refrigerant added.

THERMOCOUPLE CALIBRATION PROCEDURE

This procedure requires an accurate room temperature and an

accurate reference temperature for the thermocouple.
1. Turn the 52246 on. When the “R” starts blinking, press ENTER

and one of the arrow buttons.
2. Hold down the POWER button and the UP ARROW (▲) button

at the same time. The whole display will be shown.
3. Press ENTER twice until the number 2 is displayed and a

temperature reading.
4. This is the circuit board reference temperature. After you are sure

the temperatures have stabilized to room temperature, use the

arrows to adjust the temperature reading to match your room

temperature reference.
5. Press ENTER. The number 3 will be displayed and two

temperatures. One will be the previously adjusted room

temperature and the other will be the thermocouple reading. With

the thermocouple at a known reference temperature (place the

thermocouple [clamp] in a slush of crushed ice and water or

place the clamp on a thick copper tube with a calibrated

temperature gauge clamped to it as well.), Adjust the

temperature to match the reference temperature.
6. Press ENTER. The unit will shut-off.
7. The next time you press the POWER button, the unit will return

to normal operation.