K-Factors

K-factors represent the number of pulses the transmitter outputs per unit volume of fluid passing through the
flow meter. This number is dependent on the flow meter and the transmitter resolution setting (switch S3).
Max Machinery indicators can be adjusted to display the desired units (ccs, lbs, gallons, quarts, etc.) by
using the K-factor.

Flow meters are multi-point calibrated at the factory. The K-factor varies slightly with flow rate. A graph
of this K-factor is provided for each customer. The values shown below are typical. If the output resolution
setting is changed (via S3), the K-factors will scale proportionally:

**Maximum Count Speed of Max Machinery Indicators:
Model 120 <= 7500 Hz
Model 121 <= 20,000 Hz
If the frequency output at the application’s maximum flow rate exceeds the indicator’s count speed, then
the transmitter resolution setting (S3) will need to be adjusted to proportionally reduce the
meter/transmitter’s K-Factor.

K-Factors

New K-factor

Existing K factor

New Resolution

Existing Resolution

286-300-350 © 2002, Max Machinery, Inc.

(13)

Min

(@24 pulses/rev)

Max

(@1000 pulses/rev)

Min

(@24 pulses/rev)

Max

(@1000 pulses/rev)

213

27.6

1150.0

1.8

828

34,500

214

2.28

95

10

380

15,870

215

0.504

21.0

40

336

14,000

216

0.142

5.9

100

237

9,870

220

2.28

95.0

3

114

4,760

221

1.02

42.5

15

255

10,630

222

0.506

21.1

38

320

13,350

241

0.386

16.1

190

1,220

50,900

242

0.132

5.50

540

1,190

49,520

243

0.0417

1.74

1400

973

40,500

244

0.0139

0.579

2800

649

27,040

245

0.00396

0.165

5600

370

15,400

*typical values; may vary by up to 5% on specific flow meters.

286-3xx K-Factor Ranges (pulses/cc) with various Max flow meters*