Meriam Laminar Flow Elements User Manual

Page 15

Advertising
background image

15

Temp

°F


+0


+1


+2


+3


+4


+5


+6


+7


+8


+9

50

60

70

80

90

100

110

120

130

140

150

1.0392

1.0192

1.0000

0.9815

0.9636

0.9464

0.9298

0.9137

0.8982

0.8833

0.8688

1.0372

1.0173

0.9981

0.9797

0.9619

0.9447

0.9282

0.9122

0.8967

0.8818

0.8674

1.0352

1.0153

0.9962

0.9778

0.9601

0.943

0.9265

0.9106

0.8952

0.8803

0.8659

1.0332

1.0134

0.9944

0.976

0.9584

0.9414

0.9249

0.909

0.8937

0.8789

0.8645

1.0311

1.0115

0.9925

0.9742

0.9567

0.9397

0.9233

0.9075

0.8922

0.8774

0.8631

1.0291

1.0095

0.9906

0.9725

0.9549

0.938

0.9217

0.9059

0.8907

0.876

0.8617

1.0271

1.0076

0.9888

0.9707

0.9532

0.9364

0.9201

0.9044

0.8892

0.8745

0.8603

1.0252

1.0057

0.987

0.9689

0.9515

0.9347

0.9185

0.9028

0.8877

0.8731

0.8589

1.0232

1.0038

0.9851

0.9671

0.9498

0.9331

0.9169

0.9013

0.8862

0.8716

0.8575

1.0212

1.0019

0.9833

0.9654

0.9481

0.9314

0.9153

0.8998

0.8847

0.8702

0.8561

Troubleshooting

Problem

Probable Cause

Remedy

Low or High
DP Indication

Insufficient or improperly sized straight pipe

downstream and/or upstream of LFE.

Use 10 diameters of straight pipe

upstream and 5 diameters of straight
pipe downstream of LFE. Pipe size
should be same as LFE outlet size,
e.g. ½" NPT on LFE means 10
diameters of ½” pipe.

One or both differential pressure

connection taps plugged.

Clean or check instrument

connecting line.

If pulsation dampener is used, check stones

(Model 50 MR2 and 50 MC2 only).

If plugged, replace with matched

pair (Meriam part #A-31650).

Leak in line between LFE and readout

device.

Detect and repair.

Large-volume and/or unequal-volume

connecting lines to readout device.

Use small-volume and equal

volume connecting lines to readout
device. See Installation on page 4.

Piping reducers at inlet and/or outlet.

Do not use reducers immediately

before or after LFE.

Pulsating /

Irregular Reading

Irregular flow pattern entering LFE.

Use at least 10 diameters of straight

pipe upstream of LFE.

Leak in system line.

Detect and repair.