Meriam Laminar Flow Elements User Manual

Page 16

Advertising
background image

16

When you decide to have your LFE cleaned, please be aware of the various capabilities of calibrating

LFE’s at Meriam. Every calibration includes: 6-8 calibration points [differential pressure 1”-8”H2O], a data
sheet with raw and reduced data, calibration curve, and instruction manual. The procedure numbers are
noted in (parentheses).

Calibrations

NIST Traceable Certificate Accuracy

(% Reading)
( )

Clean per (A33555) and calibrate per (A35822) Working-Master

X

±0.72% R

( )

Clean per (A33555) and calibrate per (A35822) Master-Master

X

±0.54% R

( )

As Received calibration (A35822/A34777), clean, and calibrate [WM] X

±0.72% R

( )

As Received calibration (A35822/A34777), clean, and calibrate [MM] X

±0.54% R

( )

Accredited calibration ISO/IEC 17025 (contact Sales for details)

X

±0.64% R

( )

Nuclear or safety application (A33544)

X

±0.72% R

( )

Subcontracted per (A35352)

X

±0.50% R


Options
( )

2 additional calibration points beyond full scale- up to 12”H2O

( )

Additional calibration points beyond full scale at the following settings _____________________.

( )

Additional calibration points below 1”H2O at the following settings _______________________.

( )

Oxygen cleaning per (A50558)

( )

Calibrating the LFE with differential pressure transmitter

( )

Hydrostatic leak testing per (A33559)

( )

Pneumatic pressure test per (A70763)


The standard reference unit for flow rate is (cfm) cubic feet per minute. The additional units available are:
( )

Liters

( )

Cubic Centimeters

( )

Meters

( )

Pounds (include flowing temperature and pressure)

( )

Kilograms (include flowing temperature and pressure)

( )

Other _____________________


Time constants:
( )

Second

( )

Minute

( )

Hour


The standard reference for differential pressure is inches of water at 4°C. The additional units available are:
( )

Millimeters of water @ 4°C

( )

Centimeters of water @ 4°C

( )

Pascals

( )

Kilopascals

( )

Other _____________________


A second data sheet and calibration curve using the listed units will be included with the calibration







Advertising