mA OUT, % OUT (Percent of 4 to 20 mA)

Choose this function to provide an output from 0.00 to 24.00 milliamps. The compliance voltage is a
nominal 24 VDC to provide the driving power to your milliamp receivers.

1) Disconnect one or both input wires from

the device to be calibrated.

2) Select “mA” or “% 4 to 20mA” with slide

switch

q

.

3) Select “SOURCE” using slide switch

w

.

4) Connect the output leads of the Meriam

Model M334 to the inputs of the device
being calibrated, making sure to check
polarity. Red lead to the plus (+) input
and black lead to the minus (-) input.

The output is adjusted by turning knob

r

while the EZ-CHECK™ switch

e

is in the

“DIAL” position, or the current can be set
at the fixed points of 4.00mA (0.0%) or
20.00mA (100.0%) with switch

e

.

READ mA, READ % (Percent of 4 to 20 mA)

Choose this function to measure from 0.00 to +52.00 milliamps or -25.0 to 300.0%.

1) Open the current loop at any conve-

nient point along the signal path.

2) Select “mA” or “% 4 to 20mA” with

slide switch

q

.

3) Select “Read” using slide switch

w

.

4) Connect the red input lead (+) of the

Meriam Model M334 to the more
positive point of the break and the
black input.

Signals below 0 mA or open circuits are
indicated by 0.00 mA (-25.0%) on the
display. Signals above 52 mA are current
limited by protection circuitry.

Sourcing Milliamps

Reading Milliamp Outputs

M334

mA Loop

Calibrator

Milliamp Receiver Input

Controller

Transmitter

Computer

Logger

I/P

DCS

OUT

M334

mA Loop

Calibrator

Milliamp Output Signal

Controller

Transmitter

P/I

DCS

4.00

READ

Meriam Process Technologies

10920 Madison Ave. • Cleveland, OH 44102 Tel: 216.281.1100 • Fax: 216.281.0228 • [email protected] • www.meriam.com

Advertising