Source/Power & Measure Two Wire Transmitters
M334 Ranges and Resolution

0.00 to 24.00 mA Full Span or -25.0 to 125.0% of 4-20 mA

Accuracy

EZ Check(s) at 4 & 20mA

1

0.0 to 24.00 mA

≤ ±0.025% of Span at 4 mA & 20 mA (± 0.005mA)
≤ ± 0.05% of 24.00mA Span (± 0.01mA)

Noise

≤ ± ½ Least Significant Digit

Temperature effect

≤ ± 0.005 %/°C of FS

Loop compliance voltage

≥ 24 DCV @ 20.00mA

Loop drive capability

1200 Ω at 20 mA for 15 hours nominal; 950 Ω with Hart

Resistor enabled (334Plus)

Battery life

Source and Power measure mode ≥ 30 hrs at 12 mA

nominal; ≥ 25 hrs with backlight on (334Plus)

1

These are internal calibrated cardinal reference points and accuracy is not defined or limited by display

resolution

2-Wire Transmitter Simulation
Accuracy

Same as Source/Power & Measure

Voltage burden

≤ 2V at 20 mA

Overload/Current limit protection

24 mA nominal

Loop voltage limits

2 to 100 VDC (fuse-less protected from reverse polarity

connections)

Battery life

≥ 125 hours nominal

Voltage Read
Range and Resolution

-99.99 to +99.99 VDC Full Span (FS)

Accuracy

≤ ± 0.05 % of FS

Temperature effect

≤ ± 100 ppm/°C of FS

Input resistance

≥ 2 MΩ

Battery life

≥ 125 hours nominal

Meraim M334 Specifications (continued)

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)

Additional Information

This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration.

Test Data is available for an additional charge.

Meriam recommends a calibration interval of one year. Contact your local representative for

recalibration and repair services.