FlexMEA36

Technical Specifications FlexMEA36

Temperature compartibility

10 - 40 °C

Dimension (W x D)

31 mm x 18.5 mm

Thickness of the electrode field

12 μm

Weight

<

1

g

Base

material

Polyimide

2611

foil

Contact pads and track material

Gold

Electrode

diameter

30

μm

Interelectrode distance (centre to centre)

300 μm

Perforation

Diameters

of

the

holes

30

μm

Electrode

height Planar

Electrode

type

TiN

electrodes

(Titanium

nitride)

Isolation

type

Polyimide

2611

foil

Electrode

impedance

Approximately

50

k

Electrode

layout

grid

6

x

6

Number of recording electrodes

32

Number of reference electrodes

2 internal reference electrodes

Number

of

ground

electrodes

2

MC_Rack
Source layout in “Data Source Setup”

Configuration, 64 channels, Amplifier: FA64I/S or FA32I/S,

MEA:

FlexMEA36

(FlexMEA36.cmp).

Channel Map

“Default Map” in Layout tab of Data Display.

Cleaning

Rinse with distilled water, optional with ethanol 70 %.

Do not autoclave or sterilize FlexMEAs by heat. These MEA types are not heat-stable

and will be irreversibly damaged! Please do not use ultrasonic bath for cleaning.

Flexible Microelectrode Array with 36 electrodes for use with 32-Channel Miniature
Preamplifier MPA32I-Flex or with the ADPT-FM-32 adapter and the standard MPA32I.

28

22

11

5

29

27

21

12

6

4

30

23

20

13

10

3

31

24

17

16

9

2

32

25

18

15

8

1

26

19

14

7

28

22

11

5

29

27

21

12

6

4

30

23

20

13

10

3

31

24

17

16

9

2

32

25

18

15

8

1

26

19

14

7

2200

31000

5000

3000

350

500

18500

Dimensions in μm

Advertising