2 temperature control – Multichannel Systems MEA1060-Up Manual User Manual

Page 14

Advertising
background image

MEA Amplifier for Upright Microscopes - Manual

10

If you use a MEA-System, the isolated power supply IPS10W that is integrated in the data
acquisition computer supplies the power for operating the MEA amplifier. The power is
distributed along the MCS high grade cable. Please consider that the amplifier can only operate
properly if the supply voltage and current specifications are fulfilled, especially if you use a custom
power supply. See also the Technical Specifications and the Troubleshooting section.

5.2 Temperature

Control

The biological sample is cultured directly on the MEA. You can record without needing an
incubator, because a heating element is integrated into the amplifier and the perfusion cannula
and are both controlled by a Temperature Controller (Temperature Controller and perfusion
cannula with heating element and sensor are either included in the MEA-System or can be
ordered separately). You have several options regarding the culture chamber, such as a
semipermeable seal that guarantees stable environmental conditions.

For using the integrated heating element of the MEA amplifier, connect it with the black
cable to the Temperature Controller. The heating is active, but the cooling is passive.
Therefore, the minimum temperature is limited by the room temperature. Please refer
to the Temperature Controller Manual for more information.

Important: Please make sure that you have selected the appropriate PID coefficients
for the amplifier type (for inverted or upright microscopes) in use.

Advertising