N5102A Digital Signal Interface Module

Parallel and Parallel Interleaved Port Configuration Clock Rates

Parallel and parallel interleaved port configurations have other limiting factors for the clock and
sample rates:

logic type

Clocks per sample selection

IQ or IF digital signal type

Clocks per sample (clocks/sample) is the ratio of the clock to sample rate. For an IQ signal type, the
sample rate is reduced by the clocks per sample value when the value is greater than one. For an IF
signal or an input signal, clocks per sample is always set to one. Refer to

Table 12- 5

for the Output

mode parallel and parallel interleaved port configuration clock rates.

For Input mode, the maximum clock rate is limited by the following factors:

sample size

data type

selected filter for Pre- FIR Samples

Refer to

Table 12- 6

for the Input mode parallel and parallel interleaved port configuration clock

rates.

Table 12-5 Output Parallel and Parallel Interleaved Clock Rates

Logic Type

Signal Type

Minimum Rate

Maximum Rate

LVDS

IQ

1 x (clocks/sample) kHz

the smaller of: 100 x (clocks /sample) MHz

or

400 MHz

IF

4 kHz

400 MHz

Other

IQ

1 x (clocks/sample) kHz

the smaller of: 100 x (clocks /sample) MHz

or

150 MHz

IF

4 kHz

150 MHz

Table 12-6 Input Parallel and Parallel Interleaved Clock Rates

Logic Type

Data Type

Minimum Rate

Maximum Rate

N/A

Samples

1 kHz

100 MHz

Pre- FIR Samples

1 kHz

50

a

MHz

a.The maximum sample rate depends on the selected filter when the data rate is Pre-FIR Samples. Refer to

“Input Mode” on

page 254

for more information.