Chapter 2

55

Basic Operation

Configuring the RF Output

4.

Press

Sweep Time

to Auto.

The sweep time returns to its fastest allowable setting.

NOTE

When using an 8757D network analyzer in manual sweep mode, you must activate the signal
generator’s Manual Mode before using the

Manual Freq

softkey to control the sweep. Press

Sweep/List

>

More (2 of 3) > Manual Mode On.

Using Alternate Sweep

1.

Press the

Save

hardkey.

This opens the table editor and softkey menu for saving instrument states. Notice that the

Select

Reg

softkey is active. (For more information on saving instrument states refer to

“Using the

Instrument State Registers” on page 63

.)

2.

Turn the front panel knob until you find an available register and press

SAVE

. Remember this

saved register number. If no registers are available, you can write over an in–use register, by
pressing

Re–SAVE

.

NOTE

When you are using the PSG in a system with an 8757D network analyzer, you are limited to
using registers 1 through 9 in sequence 0 for saving and recalling states.

3.

Press

Sweep/List

>

Configure Ramp/Step Sweep

and enter new start and stop frequency values for the

ramp sweep.

4.

Press

Alternate Sweep Register

and turn the front panel knob to select the register number of the

previously saved sweep state.

5.

Press

Alternate Sweep Off On

to On.

The signal generator alternates between the original saved sweep and the current sweep. You may
need to adjust 8757D settings to effectively view both sweeps, such as setting channel 2 to
measure sensor A. Refer to

Figure 2- 7

.