Control Company 4410 DEW-POINT/WET-BULB/HUMIDITY/TEMPERATURE ALARM User Manual

Control Company Measuring instruments

Advertising
background image

SpecificationS

temperature

Range:

-40.0 to 158.0 °F / -40.0 to 70.0 °C

Resolution: 0.1°

Humidity

Range:

1.0 to 99.9%

Resolution: 0.1%

teMpeRatURe DiSpLaY MoDeS

Relative Humidity is show on the upper portion of

the display. Temperature is shown on the lower

portion of the display.
The following Temperature Display Modes are

available:
Current Temperature-

Shows the current temperature.
Dew-Point Temperature-

Shows the Dew-Point temperature. Indicated by

td

appearing to the right of the temperature display.
Wet-Bulb Temperature-

Shows the Wet-Bulb temperature. Indicated by

tw

appearing to the right of the temperature display.
Each press of the p or q button will cycle the dis-

play through the different temperature display modes.

DiSpLaYinG °f oR °c

To display the temperature readings in Fahrenheit

or Celsius, press and release the SET button. Each

press of the SET button will toggle the temperature

display between °F and °C.

RecaLL MaX/Min MeMoRY

To view the maximum and minimum readings,

press the MODE button. “MAX” appears on the

display to indicate the maximum readings recorded

are being displayed. Press the MODE button a

second time. “MIN” appears on the display to in-

dicate the minimum readings are being displayed.

If the high or low alarms have been activated, the

display will also cycle through the alarm set point

values. (See the “Alarms” section.)
The MAX and MIN temperature that is displayed is

dependant upon the temperature display mode that

has been selected (see the “Temperature Display

Modes” section). Pressing the p or q button

while displaying the MAX and MIN will cycle the

display through the different MAX/MIN temperatures

recorded (Current, Dew-Point, and Wet-Bulb).
If no buttons are pressed for 15 seconds, the

unit will return to the current temperature display

(“MAX” or “MIN” no longer appear on the display).

ReSet MaX/Min MeMoRY

While the MAX or MIN humidity and temperature

is being displayed (indicated by “MAX” or “MIN” on

the display), press the CLEAR button once. The

display will read “- - -“ and the reading will be set

to the current humidity and temperature readings.

The MAX and MIN readings may be reset indepen-

dent of each other. When reset, the humidity and

temperature readings (ambient, dew point, and wet

bulb) are reset to the current readings.

aLaRMS

High and/or low alarm points may be set for

either of the following units of measure: Humidity,

Temperature, Dew-Point Temperature, or Wet-Bulb

Temperature. A high alarm point and a low alarm

point, or both, may be set; however, the unit of

measure for the high and low alarm must be the

same. If the high alarm point is set for humidity,

the corresponding low alarm point (if used) must

also be for humidity. For example, a high alarm

point for humidity and a low alarm point for tem-

perature may not be set simultaneously.
High Alarm Set-Point (Ÿ ALARM) = An alarm will

sound when the reading rises above the set-point
Low Alarm Set-Point (” ALARM) = An alarm will

sound when the reading falls below the set-point

The user may enable/disable any or all of the

alarm set-points.

activate/De-activate the High/Low alarms

1. Press and hold the SET button until “” ALARM”

(low alarm) appears on the display.

NOTE: While setting, if no button is pressed for 15

seconds, the unit will automatically return to the current

humidity/temperature display.

2. Each press of the p or q button will toggle

the low alarm on or off.

"” ALARM OFF” indicates that the low alarm is not active.

“” ALARM ON” indicates that the low alarm is active.

3. Once the desired on/off status appears on the

display for the low alarm, press the SET button.

“Ÿ ALARM” (high alarm) will appear on the

display.
4. Each press of the p or q button will toggle

the high alarm on or off.

"Ÿ ALARM OFF” indicates that the high alarm is not active.

“Ÿ ALARM ON” indicates that the high alarm is active.

5. Once the desired on/off status appears on the

display for the high alarm, press the SET button.

The unit of measure to monitor will flash on the

display.
“%” = Humidity

°F or °C = Ambient Temperature*

td

= Dew Point Temperature*

tw

= Wet Bulb Temperature*

*The temperature unit of measure (°C or °F) will be

dependant upon the unit selected (See the “Displaying

°F or °C” section). The temperature alarm point will

automatically be switched to °F or °C based on what

has been selected.

6. Press the p or q button until the desired unit

of measure appears on the display.
7. Press the SET button, the unit will return to the

current humidity/temperature display.
When the low alarm and/or high alarm has been

activated, %

)))

will appear at the top right corner of

the display.

Setting the alarm Values

Follow these procedures once the low alarm and/or

high alarm has been activated:
1. With the current humidity/temperature on the

display, press the MODE button 3 times. The

alarm set point will appear on the display.
“” ALARM” indicates low alarm

“Ÿ ALARM” indicates high alarm

NOTE: While setting, if no button is pressed for 15

seconds, the unit will automatically return to the current

humidity/temperature display.

2. Press and release the SET button. The alarm

set point will flash on the display.
3. Press the p or q button until the desired

value appears on the display.

Press and hold the button to rapidly change the display.

4. Once the desired value appears on the display,

press the SET button. The alarm set point will no

longer flash.
5. Press the MODE button, if both alarms have

been activated, the next alarm set point will

appear. Follow steps 2 through 4 to set the next

alarm set point value. If only one alarm has

been activated, the unit will return to the current

humidity/temperature display.

alarm Sounding

The alarm will sound and LED will flash when the

reading rises above and/or falls below the alarm

points that have been activated and set. The unit

will sound an alarm for up to one (1) minute as

long as the current reading is above and/or below

the alarm set point. The LED will continue to flash

as long as the current reading is above and/or

below the alarm set point. Once the reading

returns to an “in-range” condition, the alarm will

stop sounding and the LED will stop flashing. To

silence a sounding alarm, press any button.

“in-Range” alarm Setting

If the high alarm set point is lower than the low

alarm set point, the alarm will sound when the

reading is between the two alarm set points.

This allows an alarm to be set for an “in-range”

condition instead of an “out-of-range” condition.

For example, to have the alarm sound when the

humidity is within the range of 40% to 45%, set the

Low Alarm to alarm at 45% and the High Alarm to

alarm at 40%.